Laboratory particle counter for aqueous fluids with precise flow control for contamination monitoring, size distribution analysis and component cleanliness control
The system PAMAS SVSS is a standard laboratory particle counter for pharmaceutical applications (control of parenterals, infusion solutions, ophtalmic & oral fluids) and other low-viscous liquids. The instrument may be used as “stand alone” or software controlled.
A stepper-motor-actuated precision syringe moves the sample through the laser-diode sensor at the optimum flow-rate and an accurate sample volume is achieved. Using interchangeable syringes, the PAMAS SVSS can analyse sampling volumes of different sizes, from 100 µl to 1000 ml. For small volumes, a special “Small Volume Kit” is available.
The system can be used with different light extinction sensors. The standard sensor PAMAS HCB-LD-50/50 is equipped with a flow-rate of 10 ml/min and a size range of 1 to 200 µm.
Specification | Description |
---|---|
Product Name | Pamas SVSS |
Manufacturer | Pamas |
Product Type | Particle Counter |
Measured Parameters | Particle size distribution, particle concentration |
Measuring Range | 0.3 μm to 400 μm (size), up to 1,000,000 particles/mL (concentration) |
Sample Volume | Adjustable |
Sample Flow Rate | Adjustable |
Particle Size Channels | Up to 64 channels |
Calibration | NIST-traceable calibration |
Data Storage | Internal memory and external storage options |
Communication | Ethernet, USB, RS-232, optional Wi-Fi and Bluetooth |
Software | Included Pamas Particle Explorer software |
Power Supply | 100-240 VAC, 50/60 Hz |
Display | Touchscreen LCD display |
Application | Particle counting and analysis in liquids |
Measuring range | 1–200 μm/1–400 μm according to (ISO 21501-3) |
Flow rate | Max. particle concentration: 24,000 P/ml* at 10 ml/min** |
Volumetric Laser | Class 3R, 4.2 mW |
Measurement channels | 16 or 32 freely adjustable size channels |
Sample volume | 100 μl to 1000 ml |
Pressure and Viscosity range | without pump: 0,2 - 20 bar with additional pump: 0 - 6 bar |
Measurement time | Approx. 2 min. |
Data storage capacity | NA |
Data transfer | RS-232 serial port |
Display | LCD Dispay with Keypad |
Operating temperature | 5 °C – 40 °C |
Power supply | 100V, 120V, 230V and 240VAC |
Dimensions | 400 mm x 450 mm x 280 mm (W x H x D) |
Weight | 18kg |
Classification | USP <787>, USP <788>, USP <789>, Ph.Eur. 2.9.19, BP XIII, JP XVI, KP X, IP 2.5.8 as well as raw data |
Battery backup | NA |
Calibration | ISO 21501-2, ISO 21501-3 or USP <1788> |
Pump | Automatic sample flow and volume control by wear resistant ceramic piston pump |
Software | PAMAS USP 21 CFR Part 11 Compliant |